Ingenia MR-RT Oncology Configuration

MR system for treatment planning

Product number: 781439

The comprehensive MRI solution to fit your planning -- This dedicated MRI solution is designed to integrate high-quality MR imaging into your radiation therapy planning workflows to help boost accuracy in delineating tumors and critical structures for enhanced patient care.

Contact us

Specifications Show all

Imaging
Field strength
Ingenia 1.5T & 3.0T
Bore size
70 cm
Geometric accuracy
≤ 1 mm in Ø 32 cm volume (typical)
Dedicated RT ExamCards for
Brain, head & neck, prostate and female pelvis
Flat table top
Indexing
Interloc style
Length
245 cm (including head extension)
Width
53 cm
Accessories included
3-pin Lok-Bar, adjustable coil bridge, head extension, wall mount
Other accessories
Other MR-compatible positioning and immobilization accessories, e.g. thermoplastic masks and baseplates, can be ordered directly via CIVCO
Patient transport with flat table top
FlexTrak trolley (optional)
Geometric QA analysis package
Phantom design
Grid spacing: 25 mm, Phantom holder allows placement of the phantom in three orientations (axial, coronal, sagittal)
QA analysis volume
50 x 45* x 40 cm³ (RL x AP x FH) *effective, FOV above table: 30 cm
Software
Runs from the MR console. For multi-slice 2D evaluations. Provides contour map of geometric accuracy in MR image.
Optional external laser positioning system
Type
MR compatible, green or red laser
Allowed width range
3.83 - 4.50 m
Laser phantom
Aquarius (LAP) with calibration ExamCard