Ventilation

Trilogy100

Portable Ventilator

Find similar products

Light, versatile and easy-to-use, that’s the Trilogy100 portable ventilator at its essence. With its light weight and proven technology, Trilogy100 makes invasive and non-invasive treatment less complicated for a wide range of adult and pediatric patients.

Specifications

Dimensions and weight
Dimensions and weight
Dimensions
  • 16.68 cm L x 28.45 cm W x 23.52 cm H (6.6" L x 11.2" W x 9.3" H)
Weight
  • 5 kg (11 lb)
Controls
Controls
Circuit types
  • Active with PAP; passive
Ventilation types
  • Pressure and Volume
Volume modes
  • AC; SIMV (w/PS); CV; AC-MPV
Pressure modes
  • CPAP; S; S/T; T; PC-SIMV (w/PS);PC-MVP; AVAPS-AE
IPAP
  • 4 to 50 cm H₂O
Pressure
  • 4 to 50 cm H₂O
PEEP/EPAP for active circuits
  • 0 to 25 cm H₂O
PEEP/EPAP for passive circuits
  • 4 to 50 cm H₂O
CPAP for passive circuits
  • 4 to 20 cm H₂O
PEEP for active circuits
  • 0 to 25 cm H₂O
PEEP for passive circuits
  • 4 to 25 cm H₂O
PEEP for passive circuits
  • 4 to 20 cm H₂O
Pressure support differential
  • 0 to 40 cm H₂O
Alarms
Alarms
Circuit disconnect
  • Off; 5 to 60 sec
Apnea rate
  • Off; 10 to 60 sec
High tidal volume
  • Off; 50 to 2000 ml
Low tidal volume
  • Off; 40 to 2000 ml
High minute ventilation
  • Off; 1.0 to 99 l/min
Low minute ventilations
  • Off; 0.2 to 99 l/min
High respiratory rate
  • Off; 4 to 80 BPM
Low respiratory rate
  • Off; 4 to 80 BPM
Measured patient parameters
Measured patient parameters
Respiratory rate
  • 0 to 80 BPM
Minute ventilation
  • 0 to 99 l/min
Leak rate
  • 0 to 2000 l/min
Vte/Vti
  • 0 to 2000 ml
Peak inspiratory flow
  • 0 to 2000 l/min
Peak inspiratory pressure
  • 0 to 99 cm H₂O
Mean airway pressure
  • 0 to 99 cm H₂O
% Patient triggered breaths
  • 0 to 100 %
I:E ratio
  • 9.9:1 to 1:9.9
Controls
Controls
Tidal volume
  • 50 to 2000 ml
Breath rate
  • 0 to 60 for AC mode; 1 to 60 BPM for all other modes
Inspiratory time
  • 0.3 to 5.0 sec
Rise time
  • 1 to 6
Ramp start pressure
  • 0 to 25 cm H2O for active circuits; 4 to 25 cm H2O for passive circuits; 4 to 19 cm H2O in CPAP cm H₂O
Ramp length
  • Off; 5 to 45 min
Flex
  • Off; 1 to 3
Flow trigger sensitivity
  • 1 to 9 l/min
Flow cycle
  • 10 to 90 %
Apnea rate
  • 4 to 60 BPM
Apnea rate
  • Off; 10 to 60 sec