Breathing and Respiratory care

Trilogy100

Portable Ventilator

Light, versatile and easy-to-use, that’s the Trilogy100 portable ventilator at its essence. With its light weight and proven technology, Trilogy100 makes invasive and non-invasive treatment less complicated for a wide range of adult and pediatric patients.

Shop for sleep and respiratory care products
Physician with Care Orchestrator and Trilogy

Trilogy is now connected

 

With Care Orchestrator, Trilogy is now connected to remote patient monitoring, taking home ventilator care to the next level.

 

Learn more >

Specifications Show all

Environmental
Temperature
5 – 40 °C
Relative humidity
15 - 95 %
Atmospheric pressure
110 to 60 kPa
Controls
Circuit types
Active with PAP; passive
Ventilation types
Volume control; pressure control; bilevel
Volume modes
AC; SIMV (w/PS); CV
Pressure modes
CPAP; S; S/T; T; PC-SIMV (w/PS)
IPAP
4 to 50 cm H₂O
EPAP for active circuits
0 to 25 cm H₂O
EPAP for passive circuits
4 to 50 cm H₂O
CPAP for passive circuits
4 to 20 cm H₂O
PEEP for active circuits
0 to 25 cm H₂O
PEEP for passive circuits
4 to 25 cm H₂O
Pressure support differential
0 to 30 cm H₂O
Physical
Weight
5 kg (11 lb)
Size
16.68 cm L x 28.45 cm W x 23.52 cm H (6.6" L x 11.2" W x 9.3" H)
Alarms
Circuit disconnect
Off; 10 to 60 sec
Apnea rate
Off; 10 to 60 sec
High tidal volume
Off; 50 to 2000 ml
Low tidal volume
Off; 50 to 2000 ml
High minute ventilation
Off; 1.0 to 99 l/min
Low minute ventilations
Off; 0.5 to 99 l/min
High respiratory rate
Off; 4 to 80 beats per minute
Low respiratory rate
Off; 4 to 80 beats per minute
Measured patient parameters
Respiratory rate
0 to 80 beats per minute
Minute ventilation
0 to 99 l/min
Leak rate
0 to 2000 l/min
Peak inspiratory flow
0 to 2000 l/min
Peak inspiratory pressure
0 to 99 cm H₂O
Mean airway pressure
0 to 99 cm H₂O
Percentage patient triggered breaths
0 to 100 %
I:E ratio
9.9:1 to 1:9.9
Controls
Tidal volume
50 to 2000 ml
Breath rate
1 to 60 for AC mode; 1 to 60 for all other modes
Inspiratory time
0.3 to 5.0 sec
Rise time
1 to 6
Ramp start pressure
0 to 25 cm H2O for active circuits; 4 to 25 cm H2O for passive circuits; 4 to 19 cm H2O in CPAP cm H₂O
Ramp length
Off; 5 to 45 min
Flex
Off; 1 to 3
Flow trigger sensitivity
1 to 9 l/min
Flow cycle
10 to 90 %
Apnea rate
4 to 60 beats per minute