Search terms
Light, versatile and easy-to-use, that’s the Trilogy100 portable ventilator at its essence. With its light weight and proven technology, Trilogy100 makes invasive and non-invasive treatment less complicated for a wide range of adult and pediatric patients.
Light, versatile and easy-to-use, that’s the Trilogy100 portable ventilator at its essence. With its light weight and proven technology, Trilogy100 makes invasive and non-invasive treatment less complicated for a wide range of adult and pediatric patients.
Dimensions | 16.68 cm L x 28.45 cm W x 23.52 cm H (6.6" L x 11.2" W x 9.3" H)
|
Weight | 5 kg (11 lb)
|
Circuit types | Active with PAP; passive
|
Ventilation types | Pressure and Volume
|
Volume modes | AC; SIMV (w/PS); CV; AC-MPV
|
Pressure modes | CPAP; S; S/T; T; PC-SIMV (w/PS);PC-MVP; AVAPS-AE
|
IPAP | 4 to 50
cm H₂O
|
Pressure | 4 to 50
cm H₂O
|
PEEP/EPAP for active circuits | 0 to 25
cm H₂O
|
PEEP/EPAP for passive circuits | 4 to 50
cm H₂O
|
CPAP for passive circuits | 4 to 20
cm H₂O
|
PEEP for active circuits | 0 to 25
cm H₂O
|
PEEP for passive circuits | 4 to 25
cm H₂O
|
PEEP for passive circuits | 4 to 20
cm H₂O
|
Pressure support differential | 0 to 40
cm H₂O
|
Apnea rate | Off; 10 to 60
sec
|
Circuit disconnect | Off; 5 to 60
sec
|
High tidal volume | Off; 50 to 2000
ml
|
Low tidal volume | Off; 40 to 2000
ml
|
High minute ventilation | Off; 1.0 to 99
l/min
|
Low minute ventilations | Off; 0.2 to 99
l/min
|
High respiratory rate | Off; 4 to 80 BPM
|
Low respiratory rate | Off; 4 to 80 BPM
|
Respiratory rate | 0 to 80 BPM
|
Minute ventilation | 0 to 99
l/min
|
Vte/Vti | 0 to 2000 ml
|
Leak rate | 0 to 2000
l/min
|
Peak inspiratory flow | 0 to 2000
l/min
|
Peak inspiratory pressure | 0 to 99
cm H₂O
|
Mean airway pressure | 0 to 99
cm H₂O
|
% Patient triggered breaths | 0 to 100
%
|
I:E ratio | 9.9:1 to 1:9.9
|
Tidal volume | 50 to 2000
ml
|
Breath rate | 0 to 60 for AC mode; 1 to 60 BPM for all other modes
|
Inspiratory time | 0.3 to 5.0
sec
|
Rise time | 1 to 6
|
Ramp start pressure | 0 to 25 cm H2O for active circuits; 4 to 25 cm H2O for passive circuits; 4 to 19 cm H2O in CPAP
cm H₂O
|
Ramp length | Off; 5 to 45
min
|
Flex | Off; 1 to 3
|
Flow trigger sensitivity | 1 to 9
l/min
|
Flow cycle | 10 to 90
%
|
Apnea rate | 4 to 60 BPM
|
Apnea rate | Off; 10 to 60 sec
|