Dimensions
26 (10.2) x 22 (8.6) x 14.5 (5.7) Cm/in.
Weight
3.1 (6.9) – exluding optional recorder Kg/lb
Screen
8.4 inch SVGA TFT-AM LCD display, 800 active pixels/line; resolution: 600 active lines/frame; viewing angle ±60 degrees
Stepwise deflation pressure (Adult)
Diastolic: 10-245 (1.3-32.7); Systolic: 30-270 (4.0-36.0); MAP: 20-255 (2.7-34.0) mmHg/kPa
Stepwise deflation pressure (Pediatric)
Systolic: 30-180 (4.0-24.0); Diastolic: 10-150 (1.3-20.0); MAP: 20-160 (2.7-21.3) mmHg/kPa
Stepwise deflation pressure (Neonatal)
Systolic: 30-130 (4.0-17.0); Diastolic: 10-100 (1.3-13.3); MAP: 20-120 (2.7-16.0) mmHg/kPa
NBP measurement accuracy
Maximum standard deviation: 8 mmHg; Maximum mean error: ±5mmHg
Subsequent cuff inflation (in NBP interval mode only)
The subsequent inflation pressure is determined automatically, depending on the previous measurement and patient type
SpO2 measurement range
0-100%; pulse rate measurement range 30-300 pbm. Accuracy depends on sensor
SpO2
Meets EN ISO 9919 standard
Welch Allyn* Predictive Temperature
Probe Sites: Oral, rectal, axillary; Range 26.7-43.3°C (80-110°F); Accuracy ±0.1°C (±0.2°F) in monitoring mode
Covidien Tympanic Temperature
- Range: 33°C-42°C (91.4°F-107.6°F)
- Accuracy (overall range): ±0.2°C (±0.4°F)
- Measurement response time: ≤2
Covidien** Tympanic Temperature
Equivalency modes: ear (no adjustment), oral (adjusted from ear), core (adjusted from ear), rectal (adjusted from ear)
Battery
Lithium ion: 10.8-11.1V (with a “202” form factor)
Data Output
HL7 format, via Ethernet port; serial data
Patient type
Adult, pediatric, neonatal