Power requirements
Power requirements
System input
  • 100, 120V, 220, 240VAC, 50/60Hz, 1000VA Volts, volt alternating current, hertz, volt-ampere
Monitor input
  • 100V – 240V 50/60Hz, 39W Volts, volt alternating current, hertz, volt-ampere
Workstation input
  • 100 – 240V, 50/60Hz, 550VA, 825VA Volts, volt alternating current, hertz, volt-ampere
Dimensions
Dimensions
Workstation
  • H=17", W=10", D=16.5" Inches
Control pad (optional)
  • H=2.75", W=10.5", D=8.3" Inches
Control room controller
  • H=5", W=15", D=10" Inches
Monitor
  • H=15"– 19" (depending upon the stand extension), W=15.8", D=9.7" Inches
Connection box
  • H=9.85", W=2.95", D=7.75" Inches
Ordering Information
Ordering Information
Core system
  • CORE01
Control pad
  • CPADO1
Bedrail mount
  • MNTO1
Monitor mount
  • MNTO2
  • 1. 505-0100.21, Operator's Manual, Core Integrated, 3.4X (p. 18); 505-0101.16, Operator's Manual, s5 Series FFR_iFR Option, v3.4x (p.15).
  • 2. VAL RPT, S5-Core V3.4 SW with Core Control Pad, 215-0007.02.
  • 3. Product Spec. 809480-001, 202-0407.01.
  • 4. Requirements Specification Meridian VH SW, 806000-004 (pg 84).
  • 5. An iFR cut-point of 0.89 matches best with an FFR ischemic cut-point of 0.80 with a specificity of 87.8% and sensitivity of 73.0%. (iFR Operator’s Manual 505-0101.23)
  • 6. N Engl J Med. 2012;367(11):991-1001.
  • 7. Witzenbichler B et al. Relationship Between Intravascular Ultrasound Guidance and Clinical Outcomes After Drug-Eluting Stents: The ADAPT-DES Study. Circulation 2014 Jan: 129,4;463-470
  • 8. Ahn et al. Meta-Analysis of Outcomes After Intravascular Ultrasound Guided Versus Angiography-Guided Drug-Eluting Stent Implantation in 26,503 Patients Enrolled in Three Randomized Trials and 14 Observational Studies. Am J Cardiol 2014; 113:1338-1347
  • 9. Nair A, Margolis M, Kuban B, Vince D. Automated Coronary Plaque Characterisation with Intravascular Ultrasound Backscatter: Ex Vivo Validation. EuroIntervention. 2007; 3: 113-120
  • *Safety and efficacy of VH IVUS for use in the characterization of vascular lesions and tissue types has not been established.