New

Tightrail Guardian

Motorized dilator sheath

Find similar products

Philips motorized dilator sheath - TightRail Guardian - is the latest innovation in Philips portfolio of best-in-class lead extraction tools. A shielded, motorized, dual-mode dilator sheath, TightRail Guardian has 2 modes giving physicians even more options: shielded mode as the primary operating mode, and extended mode for stalled progression. Designed with safety in mind, TightRail Guardian delivers the safety, versatility and ease of use needed to deliver the positive outcomes that patients deserve.

Features
Dual-mode selection
Dual-mode selection

Dual-mode selection

TightRail Guardian has 2 modes: shielded mode as the primary operating mode, and extended mode for stalled progression (the blades extend up to 0.4mm beyond the tip of the sheath). The dual-mode selection provides the tactical versatility to adapt to changing clinical needs.

Dual-mode selection

Dual-mode selection
TightRail Guardian has 2 modes: shielded mode as the primary operating mode, and extended mode for stalled progression (the blades extend up to 0.4mm beyond the tip of the sheath). The dual-mode selection provides the tactical versatility to adapt to changing clinical needs.

Dual-mode selection

TightRail Guardian has 2 modes: shielded mode as the primary operating mode, and extended mode for stalled progression (the blades extend up to 0.4mm beyond the tip of the sheath). The dual-mode selection provides the tactical versatility to adapt to changing clinical needs.
Click here for more information
Dual-mode selection
Dual-mode selection

Dual-mode selection

TightRail Guardian has 2 modes: shielded mode as the primary operating mode, and extended mode for stalled progression (the blades extend up to 0.4mm beyond the tip of the sheath). The dual-mode selection provides the tactical versatility to adapt to changing clinical needs.
Motorized activation
Motorized activation

Motorized activation

The ergonomically designed handle enables smooth advancement and consistency and control with the potential to reduce hand fatigue.

Motorized activation

Motorized activation
The ergonomically designed handle enables smooth advancement and consistency and control with the potential to reduce hand fatigue.

Motorized activation

The ergonomically designed handle enables smooth advancement and consistency and control with the potential to reduce hand fatigue.
Click here for more information
Motorized activation
Motorized activation

Motorized activation

The ergonomically designed handle enables smooth advancement and consistency and control with the potential to reduce hand fatigue.
Flexible and rigid shaft
Flexible and rigid shaft

Flexible and rigid shaft

The shaft has graduated flexibility and an isodiametric (smooth) inner shaft to help stay coaxial to the lead and predictably navigate the vasculature.

Flexible and rigid shaft

Flexible and rigid shaft
The shaft has graduated flexibility and an isodiametric (smooth) inner shaft to help stay coaxial to the lead and predictably navigate the vasculature.

Flexible and rigid shaft

The shaft has graduated flexibility and an isodiametric (smooth) inner shaft to help stay coaxial to the lead and predictably navigate the vasculature.
Click here for more information
Flexible and rigid shaft
Flexible and rigid shaft

Flexible and rigid shaft

The shaft has graduated flexibility and an isodiametric (smooth) inner shaft to help stay coaxial to the lead and predictably navigate the vasculature.
Bidirectional mechanism added the rotational degrees feature
Bidirectional mechanism added the rotational degrees feature

Bidirectional mechanism added the rotational degrees feature

Designed to effectively dilate commonly encountered adhesions by rotating 1080 degrees – 540 degrees clockwise and 540 degrees counterclockwise

Bidirectional mechanism added the rotational degrees feature

Bidirectional mechanism added the rotational degrees feature
Designed to effectively dilate commonly encountered adhesions by rotating 1080 degrees – 540 degrees clockwise and 540 degrees counterclockwise

Bidirectional mechanism added the rotational degrees feature

Designed to effectively dilate commonly encountered adhesions by rotating 1080 degrees – 540 degrees clockwise and 540 degrees counterclockwise
Click here for more information
Bidirectional mechanism added the rotational degrees feature
Bidirectional mechanism added the rotational degrees feature

Bidirectional mechanism added the rotational degrees feature

Designed to effectively dilate commonly encountered adhesions by rotating 1080 degrees – 540 degrees clockwise and 540 degrees counterclockwise
  • Dual-mode selection
  • Motorized activation
  • Flexible and rigid shaft
  • Bidirectional mechanism added the rotational degrees feature
See all features
Dual-mode selection
Dual-mode selection

Dual-mode selection

TightRail Guardian has 2 modes: shielded mode as the primary operating mode, and extended mode for stalled progression (the blades extend up to 0.4mm beyond the tip of the sheath). The dual-mode selection provides the tactical versatility to adapt to changing clinical needs.

Dual-mode selection

Dual-mode selection
TightRail Guardian has 2 modes: shielded mode as the primary operating mode, and extended mode for stalled progression (the blades extend up to 0.4mm beyond the tip of the sheath). The dual-mode selection provides the tactical versatility to adapt to changing clinical needs.

Dual-mode selection

TightRail Guardian has 2 modes: shielded mode as the primary operating mode, and extended mode for stalled progression (the blades extend up to 0.4mm beyond the tip of the sheath). The dual-mode selection provides the tactical versatility to adapt to changing clinical needs.
Click here for more information
Dual-mode selection
Dual-mode selection

Dual-mode selection

TightRail Guardian has 2 modes: shielded mode as the primary operating mode, and extended mode for stalled progression (the blades extend up to 0.4mm beyond the tip of the sheath). The dual-mode selection provides the tactical versatility to adapt to changing clinical needs.
Motorized activation
Motorized activation

Motorized activation

The ergonomically designed handle enables smooth advancement and consistency and control with the potential to reduce hand fatigue.

Motorized activation

Motorized activation
The ergonomically designed handle enables smooth advancement and consistency and control with the potential to reduce hand fatigue.

Motorized activation

The ergonomically designed handle enables smooth advancement and consistency and control with the potential to reduce hand fatigue.
Click here for more information
Motorized activation
Motorized activation

Motorized activation

The ergonomically designed handle enables smooth advancement and consistency and control with the potential to reduce hand fatigue.
Flexible and rigid shaft
Flexible and rigid shaft

Flexible and rigid shaft

The shaft has graduated flexibility and an isodiametric (smooth) inner shaft to help stay coaxial to the lead and predictably navigate the vasculature.

Flexible and rigid shaft

Flexible and rigid shaft
The shaft has graduated flexibility and an isodiametric (smooth) inner shaft to help stay coaxial to the lead and predictably navigate the vasculature.

Flexible and rigid shaft

The shaft has graduated flexibility and an isodiametric (smooth) inner shaft to help stay coaxial to the lead and predictably navigate the vasculature.
Click here for more information
Flexible and rigid shaft
Flexible and rigid shaft

Flexible and rigid shaft

The shaft has graduated flexibility and an isodiametric (smooth) inner shaft to help stay coaxial to the lead and predictably navigate the vasculature.
Bidirectional mechanism added the rotational degrees feature
Bidirectional mechanism added the rotational degrees feature

Bidirectional mechanism added the rotational degrees feature

Designed to effectively dilate commonly encountered adhesions by rotating 1080 degrees – 540 degrees clockwise and 540 degrees counterclockwise

Bidirectional mechanism added the rotational degrees feature

Bidirectional mechanism added the rotational degrees feature
Designed to effectively dilate commonly encountered adhesions by rotating 1080 degrees – 540 degrees clockwise and 540 degrees counterclockwise

Bidirectional mechanism added the rotational degrees feature

Designed to effectively dilate commonly encountered adhesions by rotating 1080 degrees – 540 degrees clockwise and 540 degrees counterclockwise
Click here for more information
Bidirectional mechanism added the rotational degrees feature
Bidirectional mechanism added the rotational degrees feature

Bidirectional mechanism added the rotational degrees feature

Designed to effectively dilate commonly encountered adhesions by rotating 1080 degrees – 540 degrees clockwise and 540 degrees counterclockwise

Specifications

Model Number 575-009
Model Number 575-009
TightRail Guardian inner diameter
  • 9.0 F / 0.119 in / 3.0 mm
Outer sheath inner diameter
  • 17.4 F / 0.227 in / 5.8 mm
TightRail Guardian device outer diameter
  • 16.5 F / 0.215 in / 5.5 mm
Outer sheath device outer diameter
  • 21.0 F / 0.277 in / 7.0 mm
TightRail Guardian working length
  • 18.7 in / 47.5 cm
Outer sheath working length
  • 16.0 in / 40.6 cm
Model Number 575-011
Model Number 575-011
TightRail Guardian inner diameter
  • 11.1 F / 0.145 in / 3.7 mm
Outer sheath inner diameter
  • 19.2 F / 0.253 in / 6.4 mm
TightRail Guardian device outer diameter
  • 18.3 F / 0.241 in / 6.1 mm
Outer sheath device outer diameter
  • 23.1 F / 0.303 in / 7.7 mm
TightRail Guardian working length
  • 18.7 in / 47.5 cm
Outer sheath working length
  • 16.0 in / 40.6 cm
Model Number 575-013
Model Number 575-013
TightRail Guardian inner diameter
  • 12.9 F / 0.171 in / 4.3 mm
Outer sheath inner diameter
  • 23.1 F / 0.279 in / 7.1 mm
TightRail Guardian device outer diameter
  • 20.4 F / 0.267 in / 6.8 mm
Outer sheath device outer diameter
  • 25.2 F / 0.329 in / 8.4 mm
TightRail Guardian working length
  • 18.7 in / 47.5 cm
Outer sheath working length
  • 16.0 in / 40.6 cm
Model Number 575-009
Model Number 575-009
TightRail Guardian inner diameter
  • 9.0 F / 0.119 in / 3.0 mm
Outer sheath inner diameter
  • 17.4 F / 0.227 in / 5.8 mm
Model Number 575-011
Model Number 575-011
TightRail Guardian inner diameter
  • 11.1 F / 0.145 in / 3.7 mm
Outer sheath inner diameter
  • 19.2 F / 0.253 in / 6.4 mm
See all specifications
Model Number 575-009
Model Number 575-009
TightRail Guardian inner diameter
  • 9.0 F / 0.119 in / 3.0 mm
Outer sheath inner diameter
  • 17.4 F / 0.227 in / 5.8 mm
TightRail Guardian device outer diameter
  • 16.5 F / 0.215 in / 5.5 mm
Outer sheath device outer diameter
  • 21.0 F / 0.277 in / 7.0 mm
TightRail Guardian working length
  • 18.7 in / 47.5 cm
Outer sheath working length
  • 16.0 in / 40.6 cm
Model Number 575-011
Model Number 575-011
TightRail Guardian inner diameter
  • 11.1 F / 0.145 in / 3.7 mm
Outer sheath inner diameter
  • 19.2 F / 0.253 in / 6.4 mm
TightRail Guardian device outer diameter
  • 18.3 F / 0.241 in / 6.1 mm
Outer sheath device outer diameter
  • 23.1 F / 0.303 in / 7.7 mm
TightRail Guardian working length
  • 18.7 in / 47.5 cm
Outer sheath working length
  • 16.0 in / 40.6 cm
Model Number 575-013
Model Number 575-013
TightRail Guardian inner diameter
  • 12.9 F / 0.171 in / 4.3 mm
Outer sheath inner diameter
  • 23.1 F / 0.279 in / 7.1 mm
TightRail Guardian device outer diameter
  • 20.4 F / 0.267 in / 6.8 mm
Outer sheath device outer diameter
  • 25.2 F / 0.329 in / 8.4 mm
TightRail Guardian working length
  • 18.7 in / 47.5 cm
Outer sheath working length
  • 16.0 in / 40.6 cm
  • Product availability is subject to country regulatory clearance. Please contact your local sales representative to check the availability in your country.
  • Indications: The TightRail Guardian motorized dilator sheath is intended for use in patients requiring the percutaneous dilation of tissue to facilitate removal of cardiac leads.
  • Contraindications: None known.
  • Warnings: Lead removal devices should be used at institutions with cardiothoracic surgical capabilities by physicians knowledgeable in the techniques and devices for lead removal. Complication prevention and management protocols should be in place and routinely practiced. The recommendations for lead management of the Heart Rhythm Society¹ (HRS) and European Heart Rhythm Association² (EHRA) are highly recommended for best results.
  • The safety and effectiveness of TightRail Guardian has not been established for the following: Pediatric patients (younger than 18 years old), transitional adolescent patients (between 18 and 21 years old) and the advancement of the device into the coronary sinus.
  • Lead removal devices should be used at institutions with cardiothoracic surgical capabilities by physicians knowledgeable in the techniques and devices for lead removal. Complication prevention and management protocols should be in place and routinely practiced. The recommendations for lead management of the Heart Rhythm Society¹ (HRS) and European Heart Rhythm Association² (EHRA) are highly recommended for best results.
  • 1. Kusumoto et al. 2017 HRS Expert Consensus Statement on Cardiovascular Implantable Electronic Device Lead Management and Extraction. Heart Rhythm, 2017
  • 2. Maria G Bongiorni, et al. ESC Scientific Document Group, 2018 EHRA expert consensus statement on lead extraction: recommendations on definitions, endpoints, research trial design, and data collection requirements for clinical scientific studies and registries: endorsed by APHRS/HRS/LAHRS, EP Europace, Volume 20, Issue 7, July 2018, Page 1217, https://doi.org/10.1093/europace/euy050
  • Please refer to IFU for full list of warnings.

You are about to visit a Philips global content page

Continue

You are about to visit the Philips USA website.

I understand

You are about to visit a Philips global content page

Continue

You are about to visit the Philips USA website.

I understand

Our site can best be viewed with the latest version of Microsoft Edge, Google Chrome or Firefox.