Trilogy Hospital ventilator

Trilogy EV300

Hospital ventilator

Find similar products

Transition to the future today with the next generation of Philips ventilators. The Trilogy EV300 ventilator delivers enhanced performance in noninvasive (NIV) and invasive (IV) ventilation, so patients can be treated with a single device throughout their hospital stay, regardless of changing conditions. Trilogy EV300 is designed to stay with your patients, saving your staff time and effort as patients move between hospital departments.

Features
A fit for every patient
Adjustable therapy options

Adjustable therapy options

Trilogy EV300 is user-friendly but doesn’t compromise on the advanced features of the innovative Trilogy ventilator family technology. Its advanced respiratory monitoring capabilities are designed to help you optimize your patients' therapy. Trilogy EV300 provides noninvasive (NIV) and invasive (IV) ventilation for a wide range of adult and pediatric patients. Volume and pressure modes, AVAPS-AE, SpO₂ and EtCO₂ monitoring as well as Dynamic Lung Parameters and alarms of every parameter allow for adaptable care. Flexibility of circuits enables Trilogy EV300 to be used in a wide range of patients, from infants >2.5kg to adults.
Efficient by design
Simplicity

Simplicity

Whether you are in a sub-acute or chronic care environment, Trilogy EV300 Is designed for fast setup and stress-free operation. Its simple graphical interface, 8" touchscreen, pre-configured settings, and intuitive menu navigation allow rapid settings and changes for a wide range of patients.
Easy to maintain
Maintenance and service

Maintenance and service

Trilogy EV300 includes on-site and phone support, educational training resources, extended warranties, and maintenance plans. It only requires preventative maintenance once every year, and minimal equipment is needed for calibration. With limited support and upkeep required, and a field service option, Trilogy EV300 is easy and affordable to maintain.
Proactive assistance
Tailored ventilation

Tailored ventilation

Proven innovations are designed to treat the varying needs of respiratory insufficiencies. AVAPS automatically adjusts the ventilator support to reach the desired tidal volume, while Auto EPAP proactively adjusts to the lowest effective pressure to manage the upper airway. Auto Backup delays a machine breath until your patient exhales to maximize comfort.
Ideal for any hospital setting

Cross-setting versatility

Trilogy EV300 ventilator is equipped with features to treat chronic and critical patients in any hospital setting. A modular oxygen blender and enhanced shock and vibration technology allows for smooth transport and intra-hospital transitions.
Performance you can trust

Portability for more freedom

Trilogy EV300 gives patients and the clinicians overseeing them a new level of freedom and mobility. It features a battery life of 15 hours so patients can be transitioned between hospital departments with the peace of mind that their ventilator will continue to function flawlessly. Trilogy EV300 is easy to mount onto a roll-stand.
Autonomous reactions

Digital Auto-Trak

Improve patient-ventilator synchrony with Trilogy EV300's autonomously adaptive algorithm. The Digital Auto-Trak setting provides an enhanced triggering response for patients with minimal respiratory effort while lessening the need for manual adjustments.
  • A fit for every patient
  • Efficient by design
  • Easy to maintain
  • Proactive assistance
See all features
A fit for every patient
Adjustable therapy options

Adjustable therapy options

Trilogy EV300 is user-friendly but doesn’t compromise on the advanced features of the innovative Trilogy ventilator family technology. Its advanced respiratory monitoring capabilities are designed to help you optimize your patients' therapy. Trilogy EV300 provides noninvasive (NIV) and invasive (IV) ventilation for a wide range of adult and pediatric patients. Volume and pressure modes, AVAPS-AE, SpO₂ and EtCO₂ monitoring as well as Dynamic Lung Parameters and alarms of every parameter allow for adaptable care. Flexibility of circuits enables Trilogy EV300 to be used in a wide range of patients, from infants >2.5kg to adults.
Efficient by design
Simplicity

Simplicity

Whether you are in a sub-acute or chronic care environment, Trilogy EV300 Is designed for fast setup and stress-free operation. Its simple graphical interface, 8" touchscreen, pre-configured settings, and intuitive menu navigation allow rapid settings and changes for a wide range of patients.
Easy to maintain
Maintenance and service

Maintenance and service

Trilogy EV300 includes on-site and phone support, educational training resources, extended warranties, and maintenance plans. It only requires preventative maintenance once every year, and minimal equipment is needed for calibration. With limited support and upkeep required, and a field service option, Trilogy EV300 is easy and affordable to maintain.
Proactive assistance
Tailored ventilation

Tailored ventilation

Proven innovations are designed to treat the varying needs of respiratory insufficiencies. AVAPS automatically adjusts the ventilator support to reach the desired tidal volume, while Auto EPAP proactively adjusts to the lowest effective pressure to manage the upper airway. Auto Backup delays a machine breath until your patient exhales to maximize comfort.
Ideal for any hospital setting

Cross-setting versatility

Trilogy EV300 ventilator is equipped with features to treat chronic and critical patients in any hospital setting. A modular oxygen blender and enhanced shock and vibration technology allows for smooth transport and intra-hospital transitions.
Performance you can trust

Portability for more freedom

Trilogy EV300 gives patients and the clinicians overseeing them a new level of freedom and mobility. It features a battery life of 15 hours so patients can be transitioned between hospital departments with the peace of mind that their ventilator will continue to function flawlessly. Trilogy EV300 is easy to mount onto a roll-stand.
Autonomous reactions

Digital Auto-Trak

Improve patient-ventilator synchrony with Trilogy EV300's autonomously adaptive algorithm. The Digital Auto-Trak setting provides an enhanced triggering response for patients with minimal respiratory effort while lessening the need for manual adjustments.

Specifications

Alarms
Alarms
Minute Ventilation
  • OFF, 0.2 - 30 L/min
Circuit Disconnection
  • OFF, 5 - 60 sec
Tidal Volume
  • OFF, 10 - 2000 ml
Respiratory Rate
  • OFF, 1 - 90 BPM
Inspiratory Pressure
  • 1 - 90 cmH₂O
Apnea Interval
  • 5 - 60 sec
Oxygen
Oxygen
High pressure
  • 280 to 600 kPa (41 to 87 psi)
Low flow
  • 0 to 30 l/min; maximum 10 psi
Controls
Controls
Flow trigger sensitivity
  • 0.5 - 9 l/min
Flow cycle sensitivity
  • 10% - 90% of peak flow
PEEP
  • 0 - 35 cmH₂0 for active circuits 3 - 25 cmH₂0 for passive circuits
IPAP
  • 3 - 60 cmH₂0
FiO₂
  • 21% - 100%
Triggering and cycling
  • Off, Auto-Trak, Sensitive Auto-Trak, and Flow Trigger
EPAP/CPAP
  • 3 - 25 cmH₂0
AVAPS with passive circuit
  • PSV, S/T, and A/C-PC modes only
Pressure support/control
  • 0 - 60 cmH₂0
Backup ventilation
  • ON - OFF
Inspiratory time min/max
  • 0.3 - 3.0 sec
Tidal volume
  • 35 - 2000 ml on Dual Limb and Active Flow circuits, 50 - 2000 ml on passive and active PAP circuits
Inspiratory time
  • 0.3 - 5.0 sec
Rise time
  • 0-6 sec
Breath rate
  • 0 - 80 BPM
Flow pattern
  • Square, Ramp
Standards
Standards
Particular
  • ISO 80601-2-72 Part 2-72 ISO 80601-2-12 Part 2-12 ISO 80601-2-61 Part 2-61 ISO 80601-2-55 Part 2-55
General
  • IEC 60601-1-1 Med electrical equip. Part 1-1: Gen. requirements for safety
Collateral
  • IEC 60601-1-11 Home Health Care Environment acc. to transit-operable usage
Wireless communication
  • Bluetooth v4.1, ISO/IEC 18092:2013, 21481 ed 2.0, 14443 ed 2.0, WLAN Std: IEEE 802.11 (2012) b/g/n
Ventilation modes
Ventilation modes
A/C-VC
  • Assist control (volume control)
PSV
  • Pressure support ventilation
SIMV-PC
  • Synchronized intermittent mandatory ventilation (pressure control)
CPAP
  • Continuous positive airway pressure
A/C-PC
  • Assist control (pressure control)
S/T
  • Spontaneous/timed ventilation
SIMV-VC
  • Synchronized intermittent mandatory ventilation (volume control)
Electrical
Electrical
DC input voltage
  • 12/24V 6.5A
Li-ion batteries
  • Internal/hot-swappable Li-ion batteries: 7.5 hours each nominal total run time (per IEC 80601-2-72)
Charge time for battery
  • Charge time for detachable & internal battery: from 0% to 80%: 2.5 hours from 0% to 100%: 3.5 hours
AC input voltage
  • 100V - 240V, 50/60 Hz, 1.7 - 0.6A
Environmental
Environmental
Storage temperature
  • Temperature: -25°C to 70°C Relative humidity: 5% to 93% RH, non-condensing
Transient operating temp
  • -20°C to 50°C
Operating Conditions
  • 0°C to 40°C, 5%-90% RH (non-condensing), Atm.Pressure, 62 to 106 kPa, Altitude: -1261 to 12,971 ft
Physical
Physical
Ingress protection
  • IP22: protection against finger-sized objects & dripping water when tilted up to 15 degrees.
Weight
  • 6.3 Kg (13.8 lbs) with hot-swappable battery
Size
  • 19.3 cm D x 28.6 cm W x 24.5 cm H (7.6” D x 11.25” W x 9.65” H)
Screen dimensions
  • 20.32 cm (8")
Measured and displayed patient parameters
Measured and displayed patient parameters
Dynamic compliance (Dyn C)
  • 1 to 100 ml/cmH₂O
Spontaneous triggered breaths
  • 0 to 100 %
FiO₂ with FiO₂ sensor
  • 21 to 100 %
Peak inspiratory flow PIF
  • 0 to 200 l/min
Dynamic plateau pressure
  • 0 to 90 cmH₂O
Minute ventilation (MinVent)
  • 0 to 30 l/min
Peak inspiratory pressure PIP
  • 0 to 90 cmH₂O
SpO₂ with pulse oximeter
  • 0 to 100 %
EtCO₂ with CO₂ accessory
  • 0 to 150 mmHg
Respiratory rate (RR)
  • 0 to 90 BPM
Pulse rate with pulse oximeter
  • 18 to 321 bpm
Auto-PEEP
  • 0 to 20 cmH₂O
Leak
  • 0 to 200 l/min
Dynamic resistance (Dyn R)
  • 5 to 200 cmH₂O/l/sec
I:E ratio
  • 9.9:1 to 1:9.9
Mean airway pressure
  • 0 to 90 cmH₂O
Tidal volume (Vti or Vte)
  • 0 to 2000 ml
Alarms
Alarms
Minute Ventilation
  • OFF, 0.2 - 30 L/min
Circuit Disconnection
  • OFF, 5 - 60 sec
Oxygen
Oxygen
High pressure
  • 280 to 600 kPa (41 to 87 psi)
Low flow
  • 0 to 30 l/min; maximum 10 psi
See all specifications
Alarms
Alarms
Minute Ventilation
  • OFF, 0.2 - 30 L/min
Circuit Disconnection
  • OFF, 5 - 60 sec
Tidal Volume
  • OFF, 10 - 2000 ml
Respiratory Rate
  • OFF, 1 - 90 BPM
Inspiratory Pressure
  • 1 - 90 cmH₂O
Apnea Interval
  • 5 - 60 sec
Oxygen
Oxygen
High pressure
  • 280 to 600 kPa (41 to 87 psi)
Low flow
  • 0 to 30 l/min; maximum 10 psi
Controls
Controls
Flow trigger sensitivity
  • 0.5 - 9 l/min
Flow cycle sensitivity
  • 10% - 90% of peak flow
PEEP
  • 0 - 35 cmH₂0 for active circuits 3 - 25 cmH₂0 for passive circuits
IPAP
  • 3 - 60 cmH₂0
FiO₂
  • 21% - 100%
Triggering and cycling
  • Off, Auto-Trak, Sensitive Auto-Trak, and Flow Trigger
EPAP/CPAP
  • 3 - 25 cmH₂0
AVAPS with passive circuit
  • PSV, S/T, and A/C-PC modes only
Pressure support/control
  • 0 - 60 cmH₂0
Backup ventilation
  • ON - OFF
Inspiratory time min/max
  • 0.3 - 3.0 sec
Tidal volume
  • 35 - 2000 ml on Dual Limb and Active Flow circuits, 50 - 2000 ml on passive and active PAP circuits
Inspiratory time
  • 0.3 - 5.0 sec
Rise time
  • 0-6 sec
Breath rate
  • 0 - 80 BPM
Flow pattern
  • Square, Ramp
Standards
Standards
Particular
  • ISO 80601-2-72 Part 2-72 ISO 80601-2-12 Part 2-12 ISO 80601-2-61 Part 2-61 ISO 80601-2-55 Part 2-55
General
  • IEC 60601-1-1 Med electrical equip. Part 1-1: Gen. requirements for safety
Collateral
  • IEC 60601-1-11 Home Health Care Environment acc. to transit-operable usage
Wireless communication
  • Bluetooth v4.1, ISO/IEC 18092:2013, 21481 ed 2.0, 14443 ed 2.0, WLAN Std: IEEE 802.11 (2012) b/g/n
Ventilation modes
Ventilation modes
A/C-VC
  • Assist control (volume control)
PSV
  • Pressure support ventilation
SIMV-PC
  • Synchronized intermittent mandatory ventilation (pressure control)
CPAP
  • Continuous positive airway pressure
A/C-PC
  • Assist control (pressure control)
S/T
  • Spontaneous/timed ventilation
SIMV-VC
  • Synchronized intermittent mandatory ventilation (volume control)
Electrical
Electrical
DC input voltage
  • 12/24V 6.5A
Li-ion batteries
  • Internal/hot-swappable Li-ion batteries: 7.5 hours each nominal total run time (per IEC 80601-2-72)
Charge time for battery
  • Charge time for detachable & internal battery: from 0% to 80%: 2.5 hours from 0% to 100%: 3.5 hours
AC input voltage
  • 100V - 240V, 50/60 Hz, 1.7 - 0.6A
Environmental
Environmental
Storage temperature
  • Temperature: -25°C to 70°C Relative humidity: 5% to 93% RH, non-condensing
Transient operating temp
  • -20°C to 50°C
Operating Conditions
  • 0°C to 40°C, 5%-90% RH (non-condensing), Atm.Pressure, 62 to 106 kPa, Altitude: -1261 to 12,971 ft
Physical
Physical
Ingress protection
  • IP22: protection against finger-sized objects & dripping water when tilted up to 15 degrees.
Weight
  • 6.3 Kg (13.8 lbs) with hot-swappable battery
Size
  • 19.3 cm D x 28.6 cm W x 24.5 cm H (7.6” D x 11.25” W x 9.65” H)
Screen dimensions
  • 20.32 cm (8")
Measured and displayed patient parameters
Measured and displayed patient parameters
Dynamic compliance (Dyn C)
  • 1 to 100 ml/cmH₂O
Spontaneous triggered breaths
  • 0 to 100 %
FiO₂ with FiO₂ sensor
  • 21 to 100 %
Peak inspiratory flow PIF
  • 0 to 200 l/min
Dynamic plateau pressure
  • 0 to 90 cmH₂O
Minute ventilation (MinVent)
  • 0 to 30 l/min
Peak inspiratory pressure PIP
  • 0 to 90 cmH₂O
SpO₂ with pulse oximeter
  • 0 to 100 %
EtCO₂ with CO₂ accessory
  • 0 to 150 mmHg
Respiratory rate (RR)
  • 0 to 90 BPM
Pulse rate with pulse oximeter
  • 18 to 321 bpm
Auto-PEEP
  • 0 to 20 cmH₂O
Leak
  • 0 to 200 l/min
Dynamic resistance (Dyn R)
  • 5 to 200 cmH₂O/l/sec
I:E ratio
  • 9.9:1 to 1:9.9
Mean airway pressure
  • 0 to 90 cmH₂O
Tidal volume (Vti or Vte)
  • 0 to 2000 ml
  • **Nominal run time per method in International Electrotechnical Commission (7.5 hour/battery). Detachable battery charge time 0% to 80% is 2.5 hours, internal battery charge time 0% to 100% is 3.5 hours.